AED Placement and Maintenance for Singapore Buildings

A defibrillator only works if someone can reach it, unlock it and use it within a few minutes. Most AED failures are not device failures.

The three-minute rule

Site AEDs so that any point in the building is within roughly a three-minute round trip — go, collect, return. In a tall building that means one per few floors, not one in the lobby.

Walk the route rather than measuring on a floor plan. Locked doors, lift waits and stairwells change the answer.

Where to put it

  • Somewhere always accessible — not inside an office that locks after hours.
  • High-traffic, well-known locations — lobbies, near lifts, beside fire equipment.
  • Near higher-risk areas — gyms, plant rooms, sports facilities.
  • Climate. In Singapore, an un-airconditioned cabinet in direct sun will exceed the device's rated temperature range and shorten battery and pad life. Site indoors or use a temperature-regulated cabinet.

Signage

Use the standard international AED symbol, visible from a distance and from multiple approaches. Mark the location on floor plans and fire evacuation diagrams. Tell staff at induction — a device nobody knows about is not provision.

Cabinets

Alarmed cabinets deter removal while allowing immediate access. Do not lock a publicly sited AED behind a key: the delay defeats the purpose. If theft is a real concern, an alarmed cabinet with a status window is the compromise.

Maintenance schedule

  • Weekly: check the status indicator shows ready. Most devices self-test and display a green or flashing indicator.
  • Monthly: confirm pads and battery are in date, the cabinet is intact, and spares are present.
  • After every use: replace pads, check the battery, download the event data if the device supports it.
  • Per manufacturer schedule: replace battery and pads before expiry, not after.

Assign a named owner and keep a signed log. Expiry dates should go in a diary, not in someone's memory.

Register it

Where a national or local AED registry exists, register the device and its location so emergency dispatchers can direct callers to it. An unregistered AED 100 metres from an arrest is not helping.